Reduction of Matrix Effects from Filter Effluents
Using Pall Acrodisc® MS Syringe Filters for
LCMS Sample Preparation

Gilbert E. Tumambac, Weibing Ding, and Larry Scheer, Pall Corporation, Ann Arbor, MI, USA Introduction
Liquid chromatography mass spectrometry (LCMS) is a powerful analytical technique used for various qualitative and quantitative applications. Detection is achieved in the mass spectrometry (MS) via ionization of the analytes in the chromatographic effluent.
About 75% of LCMS users utilize Electrospray Ionization (ESI) as the interface between liquid chromatography (LC) and MS. However, despite the versatility of ESI, this ionization technique suffers from a phenomenon known as the matrix effect (ME).
Matrix effect is typically attributable to components present in the sample matrix that co-elute with the compound(s) of interest.1 This competition may either suppress (ion suppression) or enhance (ion enhancement) the detection efficiency of the target analytes,which often leads to poor accuracy and precision of quantitative results.
One way to minimize the ME is by improving the method for analytical sample preparation,which typically includes sample filtration prior to analysis. If sample filtration is required, it is critical that the filters do not contribute any extractable/leachable materials that may influence the ionization efficiency of the target analytes.
This Application Note compares the cleanliness of the effluents from Pall’s Acrodisc MS syringe filters with two other commercially-available, alternative syringe filters from Competitor A (PVDF filter) and Competitor B (hydrophilic PTFE filter) by evaluating the ME due to potentially interfering extractable/leachable compounds in the LCMS analysis of three drug substances.2 Experimental Conditions
Sample solutions were prepared in triplicate by spiking three drug substances (Momet-asone Furoate, Verapamil, and Acetaminophen) individually at concentrations of 0.05 ppm,0.2 ppm, and 1 ppm into the Control (Methanol HPLC-grade) and into the effluents fromPall’s Acrodisc MS syringe filter, Competitor A’s PVDF filter, and Competitor B’s hydrophilicPTFE filter.
The effluents were obtained by passing Methanol HPLC-grade through a total of 25 filtersfrom one lot of each product using a glass syringe and collecting approximately the first 2 mL from each filter. The effluents from each filter were pooled and used as the matrix forthe three drug substances. The solutions were analyzed using a ballistic gradient ultra high performance liquid chromatographic (UHPLC)method coupled with ES-MS (electrospray-mass spectrometry) run in the Selected-Ion-Recording (SIR) mode.
The column used was an ACQUITYu UPLC BEH C18, 1.7 µm (Waters Corporation) with Mobile Phase A (LCMS-grade Methanol with 0.01% Formic Acid) and Mobile Phase B (LCMS-grade water with 0.01% Formic Acid).
ES(+) was used for the analysis of Mometasone Furoate and Verapamil, and ES(-) for Acetaminophen. The massions (parent and/or fragment) monitored for the analyses were selected based on specificity and sensitivity.
Evaluation of Matrix Effects
Evaluation of the ME was performed by comparing the area response of the target analyte spiked into the effluents from the filter (i.e. filtered Methanol) and into the unfiltered Methanol. Matrix effects were calculated as follows: Response in Unfiltered Solution (Control) where % ME = 100 indicates no matrix effect% ME > 100 indicates ion enhancement % ME < 100 indicates ion suppression Results and Discussion
The % MEs observed for each of the three drug substances in the different effluent matrices are given in Figures 1-3. The results show that, in general, the Acrodisc MS syringe filters exhibit the least influence on the detection efficiencies of the three drug substances at the concentration levels studied (% ME = 79-100%) as compared to the PVDF filter from Competitor A (% ME = 76-112%) and the hydrophilic PTFE filter from Competitor B (% ME = 13-175%) under the test conditions used in this study. This indicates that the effluentsfrom the Acrodisc MS syringe filters contain less extractable/leachable materials that may influence the detectionefficiencies of target analytes. The high degree of performance exhibited by the Acrodisc MS syringe filters canbe attributed to the low level extractable/leachable compounds from the optimized water-wettable PTFE mem-brane and the materials of construction of the device.
Comparison of % ME Observed for Mometasone Furoate at Different Concentrations in the Effluents from Pall Acrodisc MS Syringe Filters and the Filters from Competitor A and Competitor B Concentration, ppm
Pall Acrodisc MS Filter
Competitor A PVDF Filter
Competitor B PTFE Filter
Note: Dashed lines on each chart ( ) indicate no matrix effect (i.e. % ME = 100%).
Comparison of % ME Observed for Verapamil at Different Concentrations in the Effluents from Pall Acrodisc MSSyringe Filters and the Filters from Competitor A and Competitor B Concentration, ppm
Pall Acrodisc MS Filter
Competitor A PVDF Filter
Competitor B PTFE Filter
Comparison of % ME Observed for Acetaminophen at Different Concentrations in the Effluents from Pall AcrodiscMS Syringe Filters and the Filters from Competitor A and Competitor B Concentration, ppm
Pall Acrodisc MS Filter
Competitor A PVDF Filter
Competitor B PTFE Filter
Conclusions
Cleanliness of filter effluents is an important consideration for LCMS analysis. This study shows that the effluentsfrom Acrodisc MS syringe filters exhibited less ME on the detection efficiencies of the three drug substancestested as compared to those from competitive products under the same test conditions. These results indicatethat effluents from Acrodisc MS syringe filters contain less extractable/leachable materials as compared to competitive products.
Ordering Information
Acrodisc MS Syringe Filters
Part Number
References
1. Thermo Scientific. Ion suppression and matrix effects. Technical poster, 2007, www.thermo.com/us.
2. Pall Corporation Scientific and Laboratory Services, Report No. 16385GT.
Visit us on the Web at www.pall.com/lab
E-mail us at [email protected]

Pall Life Sciences
International Offices
Pall Corporation has offices and plants throughout the world in locations such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, India, Indonesia, Ireland, Italy, Japan, Korea, Malaysia, Mexico, the Netherlands, New Zealand, Norway, Poland, Puerto Rico, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, the United Kingdom, the United States, and Venezuela. Distributors in all major industrial areas of the world. To locate the Pall office or distributor nearest you, visit www.pall.com/contact.
The information provided in this literature was reviewed for accuracy at the time of publication. Product data may be subject to change without notice. For current information consult your local Pall distributor or contact Pall directly. 2012, Pall Corporation. Pall, , and Acrodisc are trademarks of Pall Corporation. ® indicates a trademark registered in the USA. Filtration.Separation.Solution. is a service mark of Pall Corporation.
uACQUITY is a trademark of Waters Corporation.

Source: http://www.pall.com/pdfs/Laboratory/12-7756_AcroMS_SmplPrep_AN.pdf

Microsoft word - document

. This Week in the Houses of the Oireachtas 24-28 September 2012 . Oireachtas Questions and Debate 17-21 September 2012 - Education: including special needs, minorities, disadvantage, literacy and numeracy, school staffing, school buildings, school patronage, curriculum - Child Services/ Children in Care: including foster care and social work services, HSE - Child Abuse/Child Protection: incl

Prix fondation pour geneve 2011 french final

Communiqué de presse 12 septembre 2011 Merck Serono s’associe à la Fondation pour Genève pour honorer le travail de recherche du Professeur Denis Duboule • Le prix 2011 de la Fondation pour Genève a été décerné au genevois Denis Duboule, spécialiste mondial de la génétique du développement • Dans le cadre de sa politique de soutien à la recherch

Copyright © 2011-2018 Health Abstracts